TH 2523

Description
TH2523 – Präzisions-Batterietester für AC-Innenwiderstand und DC-Spannung, 1 µΩ–3,5 kΩ, 100 µV–65 VDC, 1 kHz, bis 50 Messungen/s
TH 2523
TH2523 – Präzisions-Batterietester für AC-Innenwiderstand und DC-Spannung, 1 µΩ–3,5 kΩ, 100 µV–65 VDC, 1 kHz, bis 50 Messungen/s
Battery Tester
TH2523A

Description
TH2523A – Präzisions-Batterietester für AC-Innenwiderstand und Hochvolt-DC-Spannung, 1 µΩ–3,5 kΩ, 100 µV–350 VDC, 1 kHz, bis 50 Messungen/s
TH2523A
TH2523A – Präzisions-Batterietester für AC-Innenwiderstand und Hochvolt-DC-Spannung, 1 µΩ–3,5 kΩ, 100 µV–350 VDC, 1 kHz, bis 50 Messungen/s
Battery Tester

TH2523 battery tester overview

The TH2523 series is designed for combined measurement of battery AC internal resistance and DC voltage. A sinusoidal 1 kHz constant-current source excites the DUT, while the four-terminal method separates source and sense paths to minimise the influence of voltage drops in the test leads. The resistance display range extends from 1 µΩ to 3.5 kΩ. TH2523 uses two DC voltage ranges with input voltages up to 65 V, whereas TH2523A provides ranges up to 350 V for high-voltage batteries. In addition to R and V, the instrument measures impedance, reactance, inductance, capacitance, quality factor, dissipation factor and phase angle. Measurement rates up to 50 measurements per second, limit comparison, statistics, trace recording and multiple trigger and communication methods support laboratory testing and automated sequences. Operation uses a 4.3-inch colour display with 480 × 272 pixels.

Key benefits

Minimise lead resistance

The four-terminal method separates test current and voltage measurement, supporting very low resistance measurements down to the microohm range.

Measure AC internal resistance

A 1 kHz sine source with stepped constant currents measures AC resistance while reducing DC and thermoelectric contributions.

Check two quantities together

Measurement groups such as R-V combine resistance and voltage in one test; further groups cover impedance, LCR and phase parameters.

Short test cycles

Fast mode reaches up to 50 measurements per second. Front-panel, external and bus triggering coordinate the instrument with the process.

Sort directly

Nine configurable bins plus an OUT result, PASS/FAIL indication and Handler signals transfer the decision directly to a sorting or test system.

Evaluate production data

Up to 30,000 results can be analysed using mean, deviation, minimum/maximum, pass counts, Cp and Cpk.

Battery tester measurement principle

Via the LCUR, LPOT, HPOT and HCUR terminals, the TH2523 series applies a defined sinusoidal test current and measures the voltage drop on a separate path. Depending on the selected measurement group, magnitude and phase are used to determine resistance, impedance and LCR parameters. DC battery voltage can be acquired simultaneously as a separate result.

The constant-current levels of 100 mA, 10 mA, 1 mA, 100 µA and 10 µA are assigned to six resistance ranges. Automatic range selection simplifies work with changing DUTs, while HOLD keeps a selected range fixed for repeatable comparison measurements. REL, short correction and reference functions compensate for systematic influences of the measurement setup.

Important connection condition: The low test terminal is connected to earth. The DUT and test setup must therefore be checked for suitable grounding and potential relationships before connection.

Models, voltage ranges and measurement scope

ModelDC voltage rangesMaximum display valueMaximum input voltageTypical use
TH25236 V / 60 V65.000 V65 VCells, modules and low-voltage battery packs
TH2523A30 V / 300 V350.00 V350 VHigh-voltage batteries and larger packs
AC resistance rangeMaximum display valueResolution in 5-digit modeTest current
30 mΩ33.000 mΩ1 µΩ100 mA (±10%)
300 mΩ330.00 mΩ10 µΩ100 mA (±10%)
3 Ω3.3000 Ω100 µΩ10 mA (±10%)
30 Ω33.000 Ω1 mΩ1 mA (±10%)
300 Ω330.00 Ω10 mΩ100 µA (±10%)
3 kΩ3.5000 kΩ100 mΩ10 µA (±10%)

Further display ranges are 0.2 nH to 1 H for inductance, 0.001 to 9,999.9 for Q, −179.99° to 179.99° for phase angle in degrees and −3.1416 to 3.1416 in radians.

Accuracy, resolution and speed

Slow and medium modes display resistance and voltage with five digits, while fast mode uses four digits. The highest measurement rate is 50 measurements per second. Averaging from 1 to 255 measurements and a programmable delay from 0 to 60 s provide a trade-off between throughput and stable indication.

QuantityConditionBasic accuracy for 6 monthsNote
R, Z, X, LSLOW/MED, 20 °C ±5 °C30 mΩ range: ±(0.3% of reading + 0.01% of full scale)1 µΩ resolution
R, Z, X, LSLOW/MED, 20 °C ±5 °C300 mΩ to 3 kΩ: ±(0.1% of reading + 0.01% of full scale)Additional temperature coefficient ±0.05%/°C outside the reference condition
DC voltageSLOW/MED, 20 °C ±5 °C±(0.05% of reading + 0.01% of full scale)Additional temperature coefficient ±0.005%/°C outside the reference condition

The accuracy specifications require at least 30 minutes of warm-up and correctly shorted and corrected test leads. FAST mode has lower resolution and different accuracy specifications.

Measurement functions and programmable sequences

FunctionOperationBenefit
11 measurement groupsR, R-V, V, R-Q, L-Q, L-R, R-X, C-D, Z-θ°, Z-θr and R-CSuitable primary and secondary parameter combinations for battery and component tests
REL and deviationDirect display, absolute deviation or percentage deviation from a referenceFast comparison with a golden sample, nominal value or initial condition
Contact and I/V monitorChecks test-lead contact and displays signal quantities measured at the DUTReveals contact problems and makes measurement conditions visible
Trigger and delayInternal, manual, external or bus trigger; delay up to 60 sSynchronisation with a manual workstation, foot switch or test system
BIN/CompareNine limit sets with absolute or percentage evaluation, plus an OUT resultDirect classification and PASS/FAIL decision
TraceSeries with total time, interval, minimum/maximum display and automatic stop limitsMonitors changes over time without continuous operator input
StatisticsAnalysis of up to 30,000 results including mean, standard deviation, Cp and CpkEvaluation of variation and process capability across a measurement series
File managementUp to 100 instrument configurations; USB storage of results as CSV and screen captures as GIFRepeatable setups and external documentation of test results

Typical applications

Button cells and individual cells

AC internal resistance and DC voltage are combined in a compact test for fast comparison and sorting of cells.

Battery packs and e-mobility

TH2523A acquires internal resistance parameters and DC voltage on high-voltage batteries with input voltages up to 350 V.

Development and lifetime evaluation

Trace measurements record resistance and voltage changes at defined intervals; stop limits can terminate the recording automatically.

Online UPS testing

A battery can be checked in situ for voltage and AC internal resistance when potential relationships and input voltage are suitable.

Supercapacitor ESR

The 1 kHz resistance and impedance functions support evaluation of the equivalent series resistance of supercapacitors.

Contact and component resistance

Six ranges, four-terminal measurement and automatic ranging enable AC resistance checks on contacts and general components.

Interfaces, storage and automation

InterfaceAvailabilityFunction
36-pin HandlerStandardExternal trigger, handshake, BIN, PASS/FAIL and end-of-measurement signals for automated sorting and test systems
RS232CStandardSerial remote control from 9,600 to 115,200 baud
USB DeviceStandardUSBTMC instrument communication or USBCDC virtual serial port
USB HostStandardStorage of configurations, CSV measurement results and GIF screen captures on USB media
GPIBOptionalIEEE-488.2 communication and integration into existing GPIB test systems
SCPICommand setPrograms measurement function, range, trigger, comparator, statistics, trace and file functions
Trigger inputStandard; foot switch optionalStarts individual measurements from an external signal or foot switch

Installation and measurement preparation

CharacteristicRequirement or value
Test connectionFour-terminal connection via LCUR, LPOT, HPOT and HCUR; the low terminal is earthed
Warm-up timeAt least 30 minutes for specified measurement accuracy
Operating environment0 to 40 °C; up to 75% relative humidity
Dimensions without sheath215 × 87 × 335 mm (W × H × D)
Dimensions with sheath235 × 105 × 360 mm (W × H × D)
Weightapprox. 3.6 kg

For precise results, use the designated four-terminal test cables, keep contact surfaces clean, and correctly short and correct the leads before measurement. Keep test leads away from strong electromagnetic fields.

TH2523 FAQ

What is the difference between TH2523 and TH2523A?

Both models have the same AC resistance ranges. TH2523 measures DC voltage in 6 V and 60 V ranges with inputs up to 65 V; TH2523A uses 30 V and 300 V ranges with inputs up to 350 V.

Why use four-terminal measurement?

Test current flows through one lead pair while a second pair senses the voltage drop. This greatly reduces the influence of lead resistance on low resistance values.

How fast does the TH2523 series measure?

Fast mode provides up to 50 measurements per second. Averaging, trigger delay and slower modes allow a trade-off between stability and process timing.

Can the instrument be automated?

Yes. Handler, RS232C and USB Device are provided, while GPIB is optional. SCPI commands control measurement function, ranges, trigger, comparator, statistics and trace.

How are batteries sorted?

Nine bins have individual upper and lower limits; an additional OUT result identifies values outside the bins. Handler signals pass the classification to the system.

What preparation is needed for accurate measurements?

Allow at least 30 minutes of warm-up. Then correctly short the four-terminal test leads and perform SHORT and REL correction.

Can TH2523A test high-voltage batteries?

Yes, provided the input voltage does not exceed 350 V and the earthed low test terminal is compatible with the setup. Its DC voltage ranges are 30 V and 300 V.

Can degradation and lifetime be tracked?

Trace records measurements over a defined total time and at selected intervals. Stop limits can terminate the measurement series automatically.

Can supercapacitor ESR be measured?

Yes. The 1 kHz AC resistance and impedance functions support supercapacitor ESR testing within the available measurement ranges.

Technical consultation and project enquiry

Selecting TH2523 or TH2523A depends on maximum battery voltage, expected internal resistance, required measurement speed and automation needs. Please also provide the DUT, cycle time, limit logic, interfaces and intended measurement setup.