Fast, medium and slow are specified at 50, 8 and 2 readings per second respectively.
NXI-6700 – DC Voltage Measurement Module with 4 Isolated Channels
For synchronous voltage measurements on automotive electronics and electrical devices in modular test systems.
The NXI-6700-4 measures four DC voltages simultaneously in a module occupying just one slot. Three bipolar measurement ranges up to ±600 V cover low and high voltages in a compact setup; the channels are electrically isolated from one another.
The plug-in module works in an NXI-F1000 chassis or independently with a 12 V supply and LAN connection.
LAN and CAN support integration into measurement systems; SCPI, Modbus RTU and CANopen are supported.
NXI-DAQ libraries and C# and LabVIEW examples support custom voltage acquisition applications.
Personal project consultation
Do you need support selecting the right power supply? Our technical specialists will provide personal advice tailored to your application.
Select the appropriate technical configuration or access product documentation directly.
|
Description
NXI-6700-4 Multi-channel DC voltage measurement module, 4 channels, ±600 V, 0.025 % accuracy, LAN/CAN
|
Technical Data
NXI Category
Meas. & DAQ Modules
Product category
NXI Measurement & Test Systems
NXI-6700-4 DC voltage measurement module overview
The NXI-6700-4 is a four-channel module for simultaneous DC voltage acquisition. Each channel is connected across the voltage source under test using its positive and negative terminals. The ±6 V, ±60 V and ±600 V ranges cover different signal levels, with corresponding resolutions of 0.1 mV, 1 mV and 10 mV. Basic accuracy is 0.025% of reading plus 0.025% of full scale. This specification applies within one year of calibration, at 18–28 °C and up to 80% relative humidity. The manual specifies a 30-minute preparation period for precision measurements. The slim module occupies one NXI-F1000 chassis slot and can also operate independently with an external 12 V supply and LAN connection.
Key benefits
Simultaneous acquisition
Four DC voltages are measured in one module, reducing the number of separate instruments in the setup.
Three measurement ranges
±6 V, ±60 V and ±600 V accommodate different voltage levels.
Specified accuracy
Basic accuracy of 0.025% + 0.025% of full scale supports measurement planning against a defined specification.
Selectable reading rate
Three levels from 2 to 50 readings per second adapt acquisition to the task.
Separate channels
Electrical isolation between channels supports measurements at separate potentials; no isolation voltage is specified.
Modular form factor
One NXI-F1000 chassis slot or independent operation accommodates different measurement setups.
Measurement principle and technical data
Each of the four input pairs is connected across the device under test. The module measures DC voltage; it does not generate a test voltage. The maximum permissible input values must be observed when selecting ranges and wiring the system.
| Parameter | NXI-6700-4 |
|---|---|
| Channels | 4, simultaneous voltage measurement |
| DC ranges | −6 to +6 V; −60 to +60 V; −600 to +600 V |
| Resolution by range | 0.1 mV; 1 mV; 10 mV |
| Basic accuracy | 0.025% of reading + 0.025% of full scale |
| Temperature coefficient | 40 ppm/°C |
| Reading rate | Fast: 50/s; medium: 8/s; slow: 2/s |
| Supply | 12 V DC ±10% |
| Dimensions | 130.5 × 20 × 230.5 mm, including puller |
Accuracy conditions: within one year of calibration, at 18–28 °C and up to 80% relative humidity; allow 30 minutes of preparation for precision measurements.
Applications
Automotive electronics
Acquire DC voltages at four points of an automotive electronic assembly simultaneously.
Home appliances
Monitor DC voltages at several points of an appliance in a test setup.
Integrated test systems
Integrate four measurement channels into a larger acquisition system via LAN or CAN.
Interfaces and software
| Element | Function |
|---|---|
| Measurement connector | Plug-in 8-pin connector for four positive/negative input pairs |
| LAN | RJ45 connector for independent operation and PC connection |
| CAN | CAN bus communication with switchable 120 Ω termination |
| Protocols | SCPI, Modbus RTU and CANopen are supported |
| NXI DAQ & Meas. | Configuration, acquisition and recording via PC software |
| NXI-DAQ libraries | C# and LabVIEW functions and examples for custom software integration |
Installation and operation
In the NXI-F1000 chassis, the module uses the backplane connector for power and communication. Independent operation requires a suitable 12 V DC supply and LAN connection. The operating temperature is 0–40 °C; the specified environment is below 2,000 m altitude and 5–90% relative humidity without condensation. Measurement leads should be short and routed away from interference sources. External voltage must not exceed the 600 V DC measurement limit.
NXI-6700-4 FAQ
How many voltages can one module measure?
Four DC voltages can be acquired simultaneously using four separate input pairs.
Which measurement ranges are available?
The stated bipolar DC ranges are ±6 V, ±60 V and ±600 V.
Which reading rates are available?
Fast, medium and slow correspond to 50, 8 and 2 readings per second.
Can the module operate without a chassis?
Yes. Independent operation with a 12 V DC supply and LAN connection is supported.
Under which conditions does the accuracy apply?
Within one year of calibration, at 18–28 °C and up to 80% relative humidity.
Can it be used for automotive electronics?
Yes, this application is listed. The module acquires DC voltages; wiring must comply with its input limits.
Can it be integrated into test systems?
Yes, this application is listed; LAN/CAN and NXI-DAQ libraries support integration.