Selectable priority provides controlled voltage or current ramp-up according to the electrical characteristics of the device under test.
NXI-3106 – Programmable DC Power Supply Module up to 60 W
Compact supply channels for space-efficient ATE and production testing.
The 4 HP module format combines precise output control and measurement readback within a compact channel. Modules can operate independently or form part of a multi-channel NXI test system, reducing rack space compared with multiple conventional bench instruments.
Up to 1,000 sequence steps across ten files automate setpoints and individually defined dwell times.
SCPI, Modbus RTU and CANopen support integration with automated test software and higher-level control systems.
Remote sensing compensates cable voltage drops at the DUT and can be used from an output voltage of 1 V.
Personal project consultation
Do you need support selecting the right power supply? Our technical specialists will provide personal advice tailored to your application.
Select the appropriate technical configuration or access product documentation directly.
|
Description
NXI-3106-15/4 DC-Netzteilmodul, 60 W, 15 V, 4 A, CC/CV, SEQ, LAN/CAN.
|
Current [A]
4
|
Power [P]
60
|
Voltage [V]
15
|
Technical Data
NXI Category
Performance modules
Product category
NXI Measurement & Test Systems
Type of DC source
Modular Card
Standard DC Source
Benchtop Unit
|
Description
NXI-3106-30/2 DC-Netzteilmodul, 60 W, 30 V, 2 A, CC/CV, SEQ, LAN/CAN.
|
Current [A]
2
|
Power [P]
60
|
Voltage [V]
30
|
Technical Data
NXI Category
Performance modules
Product category
NXI Measurement & Test Systems
Type of DC source
Modular Card
Standard DC Source
Benchtop Unit
|
Description
NXI-3106-60/1 DC-Netzteilmodul, 60 W, 60 V, 1 A, CC/CV, SEQ, LAN/CAN.
|
Current [A]
1
|
Power [P]
60
|
Voltage [V]
60
|
Technical Data
NXI Category
Performance modules
Product category
NXI Measurement & Test Systems
Type of DC source
Modular Card
Standard DC Source
Benchtop Unit
|
Description
NXI-3106-6/12 DC power supply module, 60 W, 6 V, 12 A, CC/CV, SEQ, LAN/CAN.
|
Current [A]
12
|
Power [P]
60
|
Voltage [V]
6
|
Technical Data
NXI Category
Performance modules
Product category
NXI Measurement & Test Systems
Type of DC source
Modular Card
Standard DC Source
Benchtop Unit
NXI-3106 series overview
The NXI-3106 series comprises single-channel programmable DC power supply modules rated at 60 W. Four models cover output voltages from 6 V to 60 V and currents from 1 A to 12 A. Each module operates as a regulated DC source in CV, CC and CP modes while providing voltage and current measurement readback.
With a width of only 4 HP, the modules are suited to compact multi-channel systems for development, automated testing and production. They can operate independently through their AC input and communication interfaces or be integrated into NXI-F1080, NXI-F1030 and NXI-F1020 measurement and control chassis. A chassis configuration can provide up to 16 independently controlled supply channels.
Key benefits
Compact module format
The narrow format reduces rack space in multi-channel test systems and can replace multiple individual bench instruments.
Matched output ranges
Models rated at 6 V, 15 V, 30 V or 60 V provide a suitable voltage and current range for the application.
Fine setting resolution
Voltage resolution is 0.1 mV or 1 mV depending on the model, while current resolution is 0.1 mA or 1 mA.
Fast load recovery
A recovery time below 2 ms supports repeatable testing when the connected load changes rapidly.
Flexible channel architecture
The system can be configured from one supply channel up to a centrally controlled arrangement with 16 channels.
Measurement recording
PC software records voltage and current values, stores historical data and supports subsequent data analysis.
Programmable DC power supply operating principle
The modules provide regulated DC power for electronic assemblies, sensors, controllers and other low-voltage devices under test. In CV mode, the programmed voltage is regulated while the output current is limited to the selected maximum value. In CC mode, the unit regulates current while the output voltage is limited by the programmed voltage setting.
CP mode adds programmable power regulation to the voltage and current operating modes. Selectable CC/CV priority allows the control response to be matched to the DUT. Voltage priority supports a smooth voltage ramp, while current priority is intended for low-impedance devices or applications requiring controlled inrush current.
NXI-3106 modules operate exclusively as DC sources. Electronic load or current-sink operation is not supported.
Models and technical specifications
All variants provide 60 W through one independently controlled output channel. The maximum available current is determined by the selected model voltage range.
| Model | Voltage range | Current range | Power | Voltage resolution | Current resolution |
|---|---|---|---|---|---|
| NXI-3106-6/12 | 0–6 V | 0–12 A | 60 W | 0.1 mV | 1 mA |
| NXI-3106-15/4 | 0–15 V | 0–4 A | 60 W | 1 mV | 0.1 mA |
| NXI-3106-30/2 | 0–30 V | 0–2 A | 60 W | 1 mV | 0.1 mA |
| NXI-3106-60/1 | 0–60 V | 0–1 A | 60 W | 1 mV | 0.1 mA |
| Model | Voltage accuracy at 23 ±5 °C | Current accuracy at 23 ±5 °C | Power accuracy |
|---|---|---|---|
| NXI-3106-6/12 | 0.05% + 3 mV | 0.1% + 12 mA | 0.1% + 0.1% of full scale |
| NXI-3106-15/4 | 0.05% + 8 mV | 0.1% + 4 mA | 0.1% + 0.1% of full scale |
| NXI-3106-30/2 | 0.05% + 15 mV | 0.1% + 2 mA | 0.1% + 0.1% of full scale |
| NXI-3106-60/1 | 0.05% + 30 mV | 0.1% + 1 mA | 0.1% + 0.1% of full scale |
The stated voltage and current accuracies apply to both setting and measurement readback at 23 ±5 °C.
Dynamics and control response
The NXI-3106-6/12 and NXI-3106-15/4 models achieve rise and fall times of no more than 20 ms at no load and full load. For the NXI-3106-30/2 and NXI-3106-60/1, rise time is no more than 30 ms. Fall time is no more than 30 ms at full load and 40 ms at no load.
Following a load change from 10% to 90%, the output voltage returns to within 0.5% of rated voltage in less than 2 ms. This value describes transient recovery after a load step and should not be confused with a programmable slew rate.
| Characteristic | NXI-3106-6/12 and NXI-3106-15/4 | NXI-3106-30/2 and NXI-3106-60/1 |
|---|---|---|
| No-load rise time | ≤20 ms | ≤30 ms |
| Full-load rise time | ≤20 ms | ≤30 ms |
| No-load fall time | ≤20 ms | ≤40 ms |
| Full-load fall time | ≤20 ms | ≤30 ms |
| Transient recovery after load step | <2 ms | <2 ms |
| Regulation and output characteristic | Value |
|---|---|
| Voltage line regulation | ≤0.01% |
| Current line regulation | ≤0.05% |
| Voltage load regulation | ≤0.02% |
| Current load regulation | ≤0.05% for NXI-3106-6/12; ≤0.02% for the other models |
| Voltage ripple, 20 Hz to 20 MHz | ≤10 mV RMS / ≤60 mV peak-to-peak |
| Current ripple | ≤5 mA RMS / ≤20 mA peak-to-peak |
Operating modes and programmable functions
Constant-voltage mode
Regulates the programmed output voltage and limits current to the selected maximum value.
Constant-current mode
Regulates output current while limiting voltage according to the programmed voltage setting.
Constant-power mode
Delivers a programmed output power within the voltage, current and power limits of the selected model.
Automated setpoint profiles
Sequences control output voltage, output current and dwell time with up to 1,000 steps across ten files.
Selectable control priority
Voltage or current priority provides start-up and transition behaviour matched to the device under test.
Multi-channel control
Multiple detected modules can be selected together and switched on or off centrally within the PC software.
Typical applications
The NXI-3106 series is intended for test applications requiring several compact and independently programmable low-voltage sources.
Automated test systems
Supplies electronic assemblies during automated functional testing, with sequence operation and remote-control protocols supporting repeatable workflows.
Production and end-of-line testing
Provides defined supply voltages for production units and enables central control of multiple test channels.
Development and validation
Supports characterisation of sensors, controllers and electronic components using precisely programmed voltage and current limits.
Multi-channel assembly testing
Supplies multiple circuit sections or devices independently while reducing the required rack space.
Interfaces, software and automation
| Function | Implementation | Technical use |
|---|---|---|
| LAN | Standard, RJ45 | Network configuration, remote control and integration into PC-based test systems |
| CAN | Standard | Direct CAN communication with selectable 120 Ω termination resistor |
| SCPI | Supported | Text-based remote control within automated test programs |
| Modbus RTU | Supported | Integration into industrial control and automation environments |
| CANopen | Supported | Standardised device communication in CAN-based systems |
| Remote sense | Standard at the output | Four-wire regulation for compensating cable voltage drops from an output voltage of 1 V |
| PC software | Windows 7 or later | Hardware and channel configuration, setpoint control, data recording, history display and analysis |
The software supports device discovery, individual channel naming and central configuration of multiple modules. Recorded measurements can be stored as historical data and evaluated afterwards.
Channel integration and system configuration
An NXI-3106 module can be used as a standalone supply channel through its own single-phase AC input and integrated LAN or CAN interfaces. For multi-channel applications, it can be installed in NXI-F1080, NXI-F1030 or NXI-F1020 chassis.
A system configuration supports up to 16 channels. Each output remains independently programmable and measurable, allowing several DUTs, supply rails or assembly sections to be powered within one centrally managed test system.
Parallel or series connection of outputs to increase current, voltage or power is not specified as a standard product function. Scaling is achieved by adding independently controlled channels.
Installation and infrastructure
| Characteristic | Requirement or value |
|---|---|
| AC input | 220 V AC ±10%, single phase |
| Input frequency | 47–63 Hz |
| AC input current | ≤0.5 A |
| Dimensions with extractor handle | 130.5 × 20.0 × 230.5 mm, height × width × depth |
| Module width | 4 HP |
| Weight | Approx. 0.45 kg |
| Operating temperature | 0–40 °C |
| Storage temperature | −20 to 60 °C |
| Relative humidity | 5–90% RH, non-condensing |
| Operating altitude | <2,000 m |
| Atmospheric pressure | 80–110 kPa |
| Output-to-ground isolation | >500 V DC |
Ventilation openings must remain unobstructed during operation. The module must be switched off before connecting or disconnecting the DUT, and the protective earth connection must be correctly installed.
Protection and safety functions
Overvoltage protection
Monitors output voltage and protects the connected DUT if the configured voltage limit is exceeded.
Overcurrent protection
Limits or interrupts operation when the selected current threshold is exceeded.
Overpower protection
Monitors delivered power and prevents operation outside the permitted power range.
Overtemperature protection
Responds to excessive device temperature and protects the internal power electronics against thermal overload.
Installation, wiring and operation must be performed by appropriately qualified personnel. The device must not be operated in explosive or condensing environments.
NXI-3106 series FAQ
Which model is suitable for my application?
Select the required maximum voltage first. The available output current is then 12 A at 6 V, 4 A at 15 V, 2 A at 30 V or 1 A at 60 V.
Can the module absorb current?
No. The NXI-3106 series is a unidirectional DC source and does not provide electronic load or current-sink operation.
Can the module operate without a chassis?
Yes. The module has its own 220 V AC input together with LAN and CAN communication and supports independent operation.
How many channels can be used in one system?
An NXI system configuration supports up to 16 independently controlled channels.
Which remote-control protocols are supported?
SCPI, Modbus RTU and CANopen are supported. Physical communication is provided through LAN or CAN.
Can cable voltage drops be compensated?
Yes. The output includes remote-sense terminals for four-wire connection. Remote compensation can be used from an output voltage of 1 V.
How large can an automated sequence be?
Ten sequence files provide a combined total of up to 1,000 freely assigned steps.
What calibration interval is recommended?
Annual calibration is recommended for the NXI-3106 series.