Constant current, constant voltage, constant resistance and constant power support steady-state and dynamic testing within the respective instrument limits.
N62400S – High-Current DC Load for Ultra-Low-Voltage Testing
For fuel cell testing with high currents at very low and negative test voltages.
The N62400S series supports fuel cell testing into the negative cell-voltage region. Its largest model can draw 2,160 A at −3 V. A dedicated reverse polarity test measures the time to a defined negative end voltage, supporting investigation of cell behaviour during voltage reversal.
Up to 100 sequence files with 50 steps each store load values, slew rates, timing and acceptance limits for recurring procedures.
Remote sense measures directly at the DUT; two voltage measurement ranges also capture negative cell voltages down to −3 V.
LAN, RS232 and CAN, analogue CC programming, current monitoring and trigger connections support integration with supervisory test systems.
Personal project consultation
Do you need support selecting the right power supply? Our technical specialists will provide personal advice tailored to your application.
Select the appropriate technical configuration or access product documentation directly.
|
Description
N62406S-10-180 Low-voltage high-current electronic DC load, 600 W, 10 V, 180 A, min. operating voltage 0.2 V
|
Current [A]
180
|
Power [P]
600
|
Voltage [V]
10
|
Technical Data
Height
12
6
3
max. Current (A)
max. power (W)
max. voltage (V)
Number of channels
Single channel
Product category
DC Electronic Load
Type of DC Load
High Current
|
Description
N62412S-10-360 Low-voltage high-current electronic DC load, 1.2 kW, 10 V, 360 A, min. operating voltage 0.2 V
|
Current [A]
360
|
Power [P]
1200
|
Voltage [V]
10
|
Technical Data
Height
12
6
3
max. Current (A)
max. power (W)
max. voltage (V)
Number of channels
Single channel
Product category
DC Electronic Load
Type of DC Load
High Current
|
Description
N62418S-10-540 Low-voltage high-current electronic DC load, 1.8 kW, 10 V, 540 A, min. operating voltage 0.2 V
|
Current [A]
540
|
Power [P]
1800
|
Voltage [V]
10
|
Technical Data
Height
12
6
3
max. Current (A)
max. power (W)
max. voltage (V)
Number of channels
Single channel
Product category
DC Electronic Load
Type of DC Load
High Current
|
Description
N62424S-10-270 Low-voltage high-current electronic DC load, 2.4 kW, 10 V, 720 A, min. operating voltage 0.2 V
|
Current [A]
720
|
Power [P]
2400
|
Voltage [V]
10
|
Technical Data
Height
12
6
3
max. Current (A)
max. power (W)
max. voltage (V)
Number of channels
Single channel
Product category
DC Electronic Load
Type of DC Load
High Current
|
Description
N62436S-10-1080 Low-voltage high-current electronic DC load, 3.6 kW, 10 V, 1080 A, min. operating voltage 0.2 V
|
Current [A]
1080
|
Power [P]
3600
|
Voltage [V]
10
|
Technical Data
Height
12
6
3
max. Current (A)
max. power (W)
max. voltage (V)
Number of channels
Single channel
Product category
DC Electronic Load
Type of DC Load
High Current
|
Description
N62448S-10-1440 Low-voltage high-current electronic DC load, 3.6 kW, 10 V, 1440 A, min. operating voltage 0.2 V
|
Current [A]
1440
|
Power [P]
3600
|
Voltage [V]
10
|
Technical Data
Height
12
6
3
max. Current (A)
max. power (W)
max. voltage (V)
Number of channels
Single channel
Product category
DC Electronic Load
Type of DC Load
High Current
|
Description
N62460S-10-1800 Low-voltage high-current electronic DC load, 6 kW, 10 V, 1800 A, min. operating voltage 0.2 V
|
Current [A]
1800
|
Power [P]
6000
|
Voltage [V]
10
|
Technical Data
Height
12
6
3
max. Current (A)
max. power (W)
max. voltage (V)
Number of channels
Single channel
Product category
DC Electronic Load
Type of DC Load
High Current
|
Description
N62472S-10-2160 Low-voltage high-current electronic DC load, 7.2 kW, 10 V, 2160 A, min. operating voltage 0.2 V
|
Current [A]
2160
|
Power [P]
7200
|
Voltage [V]
10
|
Technical Data
Height
12
6
3
max. Current (A)
max. power (W)
max. voltage (V)
Number of channels
Single channel
Product category
DC Electronic Load
Type of DC Load
High Current
N62400S DC electronic load overview
The N62400S is a programmable DC electronic load for low-voltage, high-current testing with extended operation at negative cell voltages. Eight power versions provide 600 W to 7.2 kW and 180 to 2,160 A, with a maximum input voltage of 10 V. The series particularly supports fuel cell experiments in which cell voltage passes through zero and becomes negative under load. The reverse polarity test measures the interval between a trigger event or trigger voltage and a defined end voltage. CC, CV, CR and CP operation, dynamic load changes and automatically evaluated sequences supplement this function. Remote sense measures voltage directly at the DUT. Depending on the power rating, compact instruments, larger assemblies or cabinet versions are available. Model selection depends on the complete operating point, the required negative voltage range and the test system requirements.
Key benefits
Investigate voltage reversal
The RP test measures the time to a negative cell-voltage endpoint, supporting fuel cell investigation during voltage reversal.
Measure cell voltage directly
Remote sense reduces the effect of high-current cable voltage drops on voltage measurement and regulation.
Reproduce load changes
Conti, Pulse and Toggle support defined load changes with individually adjustable rise and fall slew rates.
Evaluate test steps automatically
Sequences combine load values and dwell times with voltage, current or power limit checks and PASS/FAIL results.
Determine the power point
The MPPT function examines solar sources using CV-based scans within the 10 V input limit and the current and power limits.
Optional internal resistance test
Optional DCIR measurement determines battery and supercapacitor DC internal resistance from their response to a current step.
Operating principle
The N62400S regulates loading in CC, CV, CR or CP mode. Extended low-voltage operation supports fuel cell testing at zero and negative cell voltages. The RP test applies a defined current and measures the time to a selected negative endpoint. Voltage reversal in a correctly connected DUT and reversed connecting cables are different conditions; RP testing does not replace correct wiring.
Models and operating range
| Model / version | Max. voltage | Rated current | Rated power |
|---|---|---|---|
| N62406S-10-180 | 10 V | 180 A | 600 W |
| N62412S-10-360 | 10 V | 360 A | 1,200 W |
| N62418S-10-540 | 10 V | 540 A | 1,800 W |
| N62424S (720 A version) | 10 V | 720 A | 2,400 W |
| N62436S-10-1080 | 10 V | 1,080 A | 3,600 W |
| N62448S-10-1440 | 10 V | 1,440 A | 4,800 W |
| N62460S-10-1800 | 10 V | 1,800 A | 6,000 W |
| N62472S-10-2160 | 10 V | 2,160 A | 7,200 W |
Voltage, current and power ratings must be considered together. At positive input voltages within the permitted operating envelope, P = U × I and I ≤ min(I-rated, P-rated/U).
Dynamic operation and measurement
Dynamic testing supports current, voltage, resistance and power. Conti alternates continuously between two load levels, Pulse produces a timed change after a trigger and Toggle changes level on each trigger. Slew rates and dwell times are programmable. Actual transition time depends on the selected current change, regulation and connecting cables.
| Property | Value |
|---|---|
| Voltage setting accuracy | 0.025% + 0.025% F.S. |
| Voltage readback accuracy | 0.025% + 0.025% F.S. |
| Current readback accuracy | 0.05% + 0.05% F.S. |
| CV setting resolution | 0.1 mV / 1 mV |
Accuracy values apply at 23 ±5°C. Allow 30 minutes of warm-up for precision measurements; F.S. denotes the full scale of the selected range.Measurement accuracy applies within one year after calibration.
Operating modes and test functions
| Function | Operation | Purpose |
|---|---|---|
| CC / CV / CR / CP | Steady-state current, voltage, resistance or power regulation. | Set defined load conditions. |
| RP test | Time measurement between a trigger event or trigger voltage and a negative endpoint. | Investigate fuel cell voltage reversal. |
| SEQ / Auto Test | 100 files with up to 50 steps, linking and electrical acceptance limits. | Repeatable testing with PASS/FAIL evaluation. |
| Discharge test | Constant current discharge to a defined end voltage. | Load batteries and supercapacitors. |
| Charge test | External power supply in series with the load and storage device; remote sense required. | Test CC and CC/CV charging with an external energy source. |
| MPPT | CV-mode Scan/Trace; MPP, Vmp, Imp, Voc and Isc results. | Characterise solar sources within the 10 V, current and power limits. |
| OCP / short-circuit simulation | Current steps to test the overcurrent response; short-circuit simulation as an additional test function. | Investigate a connected source’s protection behaviour. |
| DCIR / ESR, optional | Evaluate a current step and the resulting voltage change. | Determine DC internal resistance. |
Typical applications
Fuel cells for vehicle applications
Investigate cell behaviour during voltage reversal using the RP test and measure the time to a negative endpoint.
Batteries and supercapacitors
Perform defined discharge tests or investigate charging with an external power supply; DCIR is optionally available.
Solar sources up to 10 V
Determine the maximum power point using CV-based MPPT scans; open-circuit voltage, current and power must remain within instrument limits.
Automated electrical testing
Run load sequences on low-voltage sources and evaluate voltage, current or power against selected limits.
Interfaces and automation
| Feature | Function |
|---|---|
| LAN / RS232 / CAN | Digital interfaces for remote control. |
| Modbus-RTU / SCPI | Supported communication protocols; select the specific interface connection for the test system. |
| APG / trigger / current monitor | Analogue programming in CC mode, local and external triggers, and current monitoring. |
| Remote sense | Voltage measurement directly at the DUT using S+ and S−. |
| PC software | LAN-based instrument control with graphs, data logging and historical measurement data. |
Installation and infrastructure
| Item | Details |
|---|---|
| N62406S / N62412S | 220 V AC ±10%, 47–63 Hz; 132.5 × 482 × 612 mm, approximately 24 / 31 kg. |
| N62418S / N62424S | Three-phase, five-wire, 380 V AC / 50 Hz; 265 × 482 × 612 mm, approximately 43 / 50 kg. |
| N62436S / N62448S | Three-phase, five-wire, 380 V AC / 50 Hz; assemblies 487.5 or 619.5 × 482 × 612 mm. |
| N62460S / N62472S | Three-phase, five-wire, 380 V AC / 50 Hz; 22U cabinet, 600 mm wide and 800 mm deep. |
| Temperature and installation | Operation 0–40°C, full-power range 0–25°C; storage −20–60°C. Keep ventilation openings clear and size connections for the test current. |
Dimensions are H × W × D, with instrument width including handles; cabinet dimensions are listed separately.
Protection and load enable
OVP, OCP and OPP turn off the load when enabled protection thresholds are exceeded. Temperature and module overheating monitoring stop loading in the event of overheating; oscillation monitoring can shut down the load if abnormal load-current behaviour is detected. Von/Voff controls loading according to input voltage in CC mode; Von must exceed Voff. Latch mode prevents automatic restarting after unloading. RP timing and reverse connection detection serve different purposes.
Frequently asked questions
Can 10 V and 2,160 A be used simultaneously?
No. At 10 V, the N62472S-10-2160 is limited by its 7.2 kW rated power to 720 A.
What does loading at negative voltage mean?
A correctly connected fuel cell can develop a negative cell voltage under load. The N62400S supports dedicated operating points and RP testing for this condition.
What time does the RP test measure?
The interval from the selected trigger event or voltage to the defined negative cell-voltage endpoint.
Is DCIR available as standard?
The DCIR/ESR measurement function is optional.
Can the N62400S charge batteries on its own?
Charge testing requires an external power supply in series with the load and storage device, together with remote sense.
Can the series test fuel cells during voltage reversal?
Yes. The RP test investigates cell behaviour at zero or negative voltage and measures the time to the defined endpoint.
Can low-voltage tests be automated?
Yes. SEQ and Auto Test combine recurring load steps with voltage, current or power acceptance checks.