CC, CV, CR and CP provide constant current, voltage, resistance or power loading for different DC devices under test.
N62300 – Portable Programmable Electronic DC Load
For testing small power supplies, DC/DC converters and batteries in laboratories and service with programmable load sequences.
The N62300 combines a narrow, portable enclosure with dual ranges for voltage, current, resistance and power. This DC electronic load supports direct operation at the test bench and configurable, repeatable load tests on power supplies, mobile devices and small batteries.
Four SEQ groups with up to 25 steps each provide repeatable test routines with configurable timing and limits.
Von and Voff define voltages for enabling and disabling the load; short-circuit simulation provides an additional test function.
RS485 and LAN are standard interfaces. The LAN port provides PC connectivity for configuration, operation and debugging.
Personal project consultation
Do you need support selecting the right power supply? Our technical specialists will provide personal advice tailored to your application.
Select the appropriate technical configuration or access product documentation directly.
|
Description
N62320-80-30 Portable programmable DC electronic load, 200 W, 80 V, 30 A
|
Current [A]
30
|
Power [P]
200
|
Voltage [V]
80
|
Technical Data
Height
3
max. Current (A)
max. power (W)
max. voltage (V)
Number of channels
Single channel
Product category
DC Electronic Load
Type of DC Load
Benchtop Unit
Standard Load
N62300 DC electronic load overview
The N62300 is a programmable DC electronic load for testing small power supplies, DC/DC converters, consumer electronics and batteries. The N62320-80-30 model absorbs up to 200 W and supports input voltages up to 80 V and currents up to 30 A. Each operating point is constrained by voltage, current, power and minimum operating voltage simultaneously.
CC, CV, CR and CP modes provide different loading conditions. Two ranges per controlled quantity accommodate both small and larger setpoints. Front-panel keys, a rotary knob and a display for settings and measurements support local operation. Stored settings and programmable sequences simplify recurring tasks in development, production and bench repair. With a width of 72 mm and an approximate weight of 2.6 kg, the load requires little bench space. An optional 4U rack mounting arrangement accommodates up to six load channels.
Key benefits
Select the appropriate range
Two ranges per controlled quantity provide finer settings at low values, including 1 mA resolution in the 3 A current range.
Repeatable test routines
SEQ combines up to 25 steps per group with repetitions, group linking and limits for voltage, current or power.
Recall settings quickly
Four memory groups with 25 locations each keep frequently used instrument settings ready for the next device under test.
Defined switching thresholds
Von and Voff make load enabling and disabling dependent on configurable voltage values.
Investigate short-circuit behaviour
Short-circuit simulation supplements normal load points to investigate how a connected DC supply behaves under this test condition.
Record measurements on a PC
The PC application software displays measured quantities graphically and stores logged data in its history.
Operating principle and load modes
| Mode | Controlled quantity | Test task |
|---|---|---|
| CC – Constant current | Maintains the configured load current. | Load power supplies and batteries at a defined current. |
| CV – Constant voltage | Regulates the input voltage by adjusting the current drawn. | Investigate DC sources at a specified voltage point. |
| CR – Constant resistance | Adjusts the voltage-to-current ratio according to the resistance setpoint. | Emulate a resistive DC load. |
| CP – Constant power | Adjusts load current to input voltage to absorb the configured power. | Load a device at a specified power level. |
Model and usable operating range
| Model | Input voltage | Input current | Absorbed power | Minimum operating voltage |
|---|---|---|---|---|
| N62320-80-30 | 0–80 V | 0–30 A | 0–200 W | 0.5 V at 3 A; 1.5 V at 30 A |
Consider voltage, current and power together: P = U × I applies. For U > 0, within the permitted operating limits, I_max = min(30 A, 200 W/U). At 80 V, the maximum current is therefore 2.5 A. The minimum operating voltage also applies; a nominal range starting at 0 V does not mean full current can be drawn at 0 V.
Ranges, resolution and accuracy
| Quantity | Low range | High range | Setting resolution |
|---|---|---|---|
| Voltage | 0–10 V | 0–80 V | 1 mV / 10 mV |
| Current | 0–3 A | 0–30 A | 1 mA / 10 mA |
| Power | Up to 20 W | Up to 200 W | 1 mW / 10 mW |
| Resistance | Up to 999.9 Ω | Up to 9,999 Ω | 0.1 Ω / 1 Ω |
Voltage setting accuracy and voltage readback accuracy are each 0.05% + 0.1% F.S.; current setting accuracy and current readback accuracy are each 0.1% + 0.1% F.S. at 23 ±5 °C. F.S. denotes the respective full-scale value. Voltage and current inputs must exceed 10% F.S. to achieve the specified accuracy. Allow a 30-minute warm-up before measurement; measurement accuracy is valid within one year after calibration.
Communication response time is no more than 5 ms. This value describes communication and does not specify current rise time or control bandwidth.
Sequences and programmable functions
| Function | Scope | Benefit |
|---|---|---|
| SEQ | Four groups A–D with up to 25 steps each; configurable cycle count and group linking. | Combine multiple load points into a repeatable routine. |
| Step parameters | CC, CV, CR or CP, setpoint, delay and optional lower and upper limits for voltage, current or power. | Define loading, duration and test limits for each step. |
| Timing units | Milliseconds, seconds, minutes and hours. | Configure both short and longer holding periods. |
| Save/Recall | Four groups with 25 memory locations each. | Reuse prepared settings directly. |
| Von/Voff and Short | Configurable load-on and load-off voltages plus short-circuit simulation. | Include voltage thresholds and a simulated short circuit in tests. |
Typical applications
AC/DC power supplies and DC/DC converters
Connect the load to the DC output. CC, CR and CP provide defined loading of small power supply modules.
Consumer electronics and bench repair
Load supplies for smartphones and mobile devices at the bench and compare measurements using repeatedly recalled settings.
Small batteries for drones and portable devices
Load batteries as DC sources within the voltage, current and power limits. CC or CP defines the loading; Voff provides a configurable cutoff voltage.
Production and end-of-line testing
Structure recurring tests on small DC supplies using stored settings and SEQ steps with defined limits.
Interfaces, software and automation
| Equipment | Function and scope |
|---|---|
| RS485 | Standard serial communication interface. |
| LAN | Standard Ethernet port for PC connection, configuration and debugging. |
| Protocols | Modbus-RTU and SCPI; the specific mapping to the communication path must be considered during integration. |
| PC application software | Operation of CC, CV, CR, CP and SEQ, on/off control, short-circuit simulation, graphical display and data logging. |
| Communication monitoring | Software settings for timeouts, retries and behaviour following a connection loss. |
Enclosure, rack mounting and operating conditions
| Item | Specification |
|---|---|
| Dimensions and weight | 131.5 × 72 × 325 mm (H × W × D); approximately 2.6 kg net. |
| Rack mounting | Optional 4U arrangement for up to six load channels. |
| AC supply | 110/220 V AC with voltage selector, 50–60 Hz; observe the rear-panel rating and selector position. |
| Temperature | Operation 0–40 °C; full power up to 30 °C, derating above this temperature. Storage −20–60 °C. |
| Environment | Altitude below 2,000 m; relative humidity 5–90%, non-condensing. |
| Connections and cooling | Provide protective earthing, size load leads for the test current and keep ventilation openings clear. |
Protection functions
The N62300 provides overcurrent, overvoltage, overpower and overtemperature protection (OCP, OVP, OPP, OTP). OCP, OVP and OPP values are configurable. A triggered protection event is identified by its type on the error display. Connect the load with correct polarity and operate it within its electrical limits.
Frequently asked questions about the N62300
Can I use 80 V and 30 A simultaneously?
No. At 80 V, the 200 W power rating limits current to 2.5 A. Temperature derating must also be considered at higher ambient temperatures.
How low can the input voltage be?
Minimum operating voltage is 0.5 V at 3 A or 1.5 V at 30 A. The permissible current therefore also depends on input voltage.
How extensive can sequences be?
Four groups with up to 25 steps each are available. Cycle count, group linking, timing and step-specific limits are configurable.
Which PC connections are available?
RS485 and LAN are provided. LAN connects the PC application software for configuration, operation and debugging.
Can the load be rack mounted?
Yes. An optional 4U rack mounting arrangement accommodates up to six load channels.
Can I test AC/DC power supplies and DC/DC converters?
Yes. The N62300 loads their DC outputs in CC, CV, CR or CP within its electrical limits.
Is the series suitable for mobile electronics?
Yes. It is intended for tests on consumer electronics such as smartphones and mobile devices; stored settings support recurring load tests.
Can I load small drone batteries?
Yes. Small batteries can be load tested within 80 V, 30 A and 200 W, with all limits applying simultaneously. CC or CP defines the load, while Voff sets the cutoff voltage.