Skip to main content Skip to search Skip to main navigation

Filter products

ART. NO.
Description
Current [A]
Power [P]
Voltage [V]
N6136-100-05

Description
N6136-100-05 Multi-channel programmable DC electronic load, 36 channels, 50 W per channel, 100 V, 5 A, 3U
Current [A]
5
Power [P]
50
Voltage [V]
100
N6136-100-05
N6136-100-05 Multi-channel programmable DC electronic load, 36 channels, 50 W per channel, 100 V, 5 A, 3U
5
50
100
36/48 CH DC Load

Technical Data

max. Current (A)

max. power (W)

max. voltage (V)

Number of channels

Multichannel

Product category

DC Electronic Load

Type of DC Load

Multi-channel

N6148-100-05

Description
N6148-100-05 Multi-channel programmable DC electronic load, 48 channels, 50 W per channel, 100 V, 5 A, 3U
Current [A]
5
Power [P]
50
Voltage [V]
100
N6148-100-05
N6148-100-05 Multi-channel programmable DC electronic load, 48 channels, 50 W per channel, 100 V, 5 A, 3U
5
50
100
36/48 CH DC Load

Technical Data

max. Current (A)

max. power (W)

max. voltage (V)

Number of channels

Multichannel

Product category

DC Electronic Load

Type of DC Load

Multi-channel

N6148 multi-channel DC electronic load overview

The N6148 series is a programmable DC electronic load for development, production, ageing tests and quality control of DC/DC converters and power supplies. It draws energy from the device under test and regulates the load in CC, CR or CP mode. The series comprises the N6136-100-05 with 36 channels and the N6148-100-05 with 48 channels. Both versions provide up to 100 V, 5 A and 50 W per channel; the power limit must be considered when selecting the operating point.

Load channels are grouped in modules with six channels each. The 19-inch, 3U enclosure suits test systems requiring many channels. An LCD, rotary knob and keys provide local operation. PC software enables users to configure selected channels together and switch them on or off. LAN and RS485 support integration into automated test setups.

Key benefits

Many test points in a compact space

36 or 48 channels in one instrument reduce space requirements compared with an equivalent number of small standalone loads.

Fewer repeated entries

The PC software applies operating modes and parameters to selected channel groups and provides Batch On and Batch Off.

Channel status at a glance

All View displays operating mode, voltage, current and channel status for quick checks at the instrument.

Channel-specific limits

Voltage, current and power protection parameters can be adjusted for each channel to suit the test task.

Operating principle and load modes

ModeOperationUse
CC – constant currentRegulates the configured load current as input voltage changes.Defined current loading of supplies and current paths.
CR – constant resistanceLoad current changes with input voltage and the configured resistance.Reproducing a resistive load.
CP – constant powerAdjusts current to the input voltage to maintain the configured power.Testing at a specified power consumption.

All modes operate within the permissible voltage, current and power limits of the individual channel.

Models and operating range

ModelChannelsVoltage per channelCurrent per channelPower per channel
N6136-100-05360–100 V0–5 A0–50 W
N6148-100-05480–100 V0–5 A0–50 W

Select the operating point for each channel: Voltage and current are related to power through P = U × I. 100 V and 5 A cannot be used simultaneously. At 100 V, the 50 W power rating gives a calculated current ceiling of 0.5 A. The channel count describes separate load channels, not a combined high-power load.

Setting ranges and resolution

QuantityRangeResolution
CC current setting0–5 A1 mA
CR resistance setting1 Ω–2 kΩ0.1 Ω
CP power setting0–50 W10 mW
Voltage measurement0–100 V10 mV

Resolution describes the setting or measurement increment and is distinct from accuracy. The communication response time is no more than 10 ms; it does not specify current rise time or settling time.

OCP testing and voltage-dependent load activation

During OCP testing, the load increases current in steps in CC mode while monitoring DUT voltage. If the voltage falls below the configured end voltage, the current load value is recorded as the result and the load input is switched off. Otherwise, current continues to increase up to the configured maximum.

LatchCurrent sinkingBehaviour after a voltage drop
DisabledStarts above Von.Stops below Von and can resume when the voltage exceeds Von again.
EnabledStarts above Von.Stops below Voff. There is no automatic restart even if the voltage exceeds Von again.

The OCP test evaluates the DUT overcurrent protection. The electronic load has a separate overcurrent monitoring function for its own protection.

Typical applications

Wiring harnesses and connectors

Apply adjustable DC loads to electrical current paths in a powered test setup. Select current and power for each channel.

Interfaces, software and automation

EquipmentFunction
LAN – built inEthernet connection to the PC; device discovery and hardware configuration in the application software.
RS485 – built inSerial remote control through terminals A, B and G. A USB-to-RS485 adapter can provide the PC connection.
SCPI and ModbusSupported communication protocols for system integration.
PC application softwareChannel naming, individual and group CC/CR/CP configuration, protection parameters and Batch On/Off.
Communication lossThe software offers a choice between “No Action” and “Device OFF”.
Local operationLCD, rotary knob, keys and All View for displaying channel status.

For LAN communication, the PC and instrument are configured on the same network segment with different IP addresses. Group operation simplifies the configuration of multi-channel tests.

Installation and infrastructure

FeatureConfiguration or requirement
Form factor19-inch, 3U
AC supplySingle phase, 220 V AC ±10%, 47–63 Hz; observe the voltage marking on the instrument.
Operating temperature0–40 °C
EnvironmentAltitude below 2,000 m; 5–90% relative humidity, non-condensing; 80–110 kPa.
DUT connectionSelect polarity, wire cross-section and cable length to suit the current and test setup.
Measurement preparationAllow 30 minutes of warm-up time; an annual calibration interval is recommended.

Protection functions

OCP, OVP, OPP and OTP monitor overcurrent, overvoltage, overpower and overtemperature. Overcurrent and overvoltage events are indicated on the instrument and in the PC software. Overpower protection disconnects the load input. In an overtemperature condition, the load switches off and the fan runs at full speed to cool the instrument.

Voltage, current and power limits are configured by channel. These protection functions and the selectable response to communication loss should be considered when designing the test procedure.

Frequently asked questions about the N6148 series

Which channel versions are available?

The N6136-100-05 has 36 channels and the N6148-100-05 has 48 channels. Both provide a maximum of 100 V, 5 A and 50 W per channel.

Can 100 V and 5 A be used simultaneously?

No. This would require 500 W per channel. The power rating is 50 W, giving a calculated current ceiling of 0.5 A at 100 V.

Which operating modes can be selected?

CC, CR and CP provide constant current, constant resistance and constant power operation. Modes and parameters are selected by channel.

Does the load restart automatically after a voltage dip?

With the Von latch enabled, it remains off after the voltage falls below Voff, even if voltage rises above Von again. Without the latch, current sinking follows the Von threshold.

Can several channels be operated together?

Yes. The PC software can configure selected channels together and switch them using Batch On or Batch Off.

Can the overcurrent protection of a DC/DC converter be tested?

Yes. The OCP test increases load current in steps and records the current when voltage falls below the configured end voltage. Test limits must suit both the converter and the load channel.

Can the series be used for power supply ageing tests?

Yes. The channels provide defined CC, CR or CP loads for ageing and quality tests. Electrical limits and permissible environmental conditions must be observed.

Can the N6148 load wiring harnesses and electrical distribution centres?

Yes. The series is intended for testing wiring harnesses, connectors, fuses, relays and BECs. In a powered test setup, it provides the electronic load within the channel limits.

Technical consultation and project enquiry

Planning a multi-channel test bench with the N6148 series? Tell us the required channel count, voltage, current and power at each test point, together with operating modes, interfaces and test procedure. We can support instrument selection and integration into your test setup.

Discuss your N6148 project with ET System

Get in touch with us

+49 (0) 6205 - 3948-0 info@et-system.de

Mon - Thu: 8:00 AM - 5:00 PM
Fri: 8:00 AM - 3:00 PM

Contact Us Directly

+49 (0) 6205 - 3948-0 info@et-system.de

Mo - Do von 8 - 17 Uhr
Fr von 8 - 15 Uhr

Contact Us Directly