N3225-2.5K-M010

Description
N3225-2.5K-M010 Programmable high-voltage DC power supply, 25 W, ±2500 V, 10 mA, 2U, LAN/RS232
Current [A]
0,01
Power [P]
25
Voltage [V]
2500
N3225-2.5K-M010
N3225-2.5K-M010 Programmable high-voltage DC power supply, 25 W, ±2500 V, 10 mA, 2U, LAN/RS232
0,01
25
2500
High-voltage DC Source

Technical Data

Product category

DC Power Supply

Type of DC source

Standard DC Source

High Voltage

Benchtop Unit

N3225-5K-M005

Description
N3225-5K-M005 Programmable high-voltage DC power supply, 25 W, ±5000 V, 5 mA, 2U, LAN/RS232
Current [A]
0,005
Power [P]
25
Voltage [V]
5000
N3225-5K-M005
N3225-5K-M005 Programmable high-voltage DC power supply, 25 W, ±5000 V, 5 mA, 2U, LAN/RS232
0,005
25
5000
High-voltage DC Source

Technical Data

Product category

DC Power Supply

Type of DC source

Standard DC Source

High Voltage

Benchtop Unit

N3225-10K-M002

Description
N3225-10K-M002 Programmable high-voltage DC power supply, 20 W, ±10000 V, 2 mA, 2U, LAN/RS232
Current [A]
0,002
Power [P]
20
Voltage [V]
10000
N3225-10K-M002
N3225-10K-M002 Programmable high-voltage DC power supply, 20 W, ±10000 V, 2 mA, 2U, LAN/RS232
0,002
20
10000
High-voltage DC Source

Technical Data

Product category

DC Power Supply

Type of DC source

Standard DC Source

High Voltage

Benchtop Unit

N3200 high-voltage DC power supply overview

The N3200 programmable high-voltage DC power supply is designed for tests that combine high voltage with low current. Three single-channel models cover positive and negative voltage ranges up to 2.5 kV, 5 kV or 10 kV. The two lower voltage classes deliver up to 25 W, while the 10 kV version provides up to 20 W. Applications include breakdown testing of high-voltage components, insulation testing, leakage current measurements and investigations of high-resistivity materials. Voltage can be adjusted from 0 V in the selected polarity. Source mode provides defined operating points, while sweep mode changes voltage in steps. Operation is available locally through the LCD, keys and rotary knob, or remotely through the communication and analogue connections. Rear high-voltage terminals and the compact enclosure support integration into laboratory or automated high-voltage test systems.

Key benefits

Set small increments

Voltage setting and readback resolution is 0.1 V up to 5 kV and 1 V on the 10 kV model. Current resolution is 0.1 µA across all models.

Low output ripple

Ripple is at most 3 mV RMS up to 5 kV and at most 10 mV RMS at 10 kV, measured with a filter from 20 Hz to 20 MHz. This supports sensitive measurement setups.

Select output polarity

The Output Reverse function selects output polarity. Positive and negative test voltages can be set within the voltage range of each model.

Change test voltage in steps

Sweep mode combines voltage limits, increment and time interval. Defined voltage sequences can be reused for component investigations.

Connect analogue measurement chains

VSET for voltage programming and VMON and IMON for monitoring enable integration with analogue control and data acquisition systems.

Document measurement trends

The PC software displays voltage and current traces and stores readings. Export and historical data viewing support subsequent analysis.

Operating principle and output polarity

Source mode provides settings for output voltage and current parameters. Output Reverse selects the positive or negative output direction; the ± rating describes these selectable voltage ranges of a single channel. In sweep mode, I-Out limits current while the output voltage changes in predefined steps. Adjustable voltage and current trip thresholds complement these settings.

Models and output ratings

ModelAdjustable output voltageMax. currentRated powerChannels
N3225-2.5K-M010−2,500 to 0 V / 0 to +2,500 V10 mA25 W1
N3225-5K-M005−5,000 to 0 V / 0 to +5,000 V5 mA25 W1
N3225-10K-M002−10,000 to 0 V / 0 to +10,000 V2 mA20 W1

Select voltage and current together for each model: the highest voltage of 10 kV is available from the N3225-10K-M002 at up to 2 mA. The 10 mA rating of the N3225-2.5K-M010 applies within its voltage range up to 2.5 kV.

Resolution, ripple and dynamic behaviour

ParameterN3225-2.5K-M010N3225-5K-M005N3225-10K-M002
Voltage resolution, setting / readback0.1 V0.1 V1 V
Current resolution, setting / readback0.1 µA0.1 µA0.1 µA
Ripple, 20 Hz–20 MHz, with filter≤15 mV peak-to-peak / ≤3 mV RMS≤30 mV peak-to-peak / ≤3 mV RMS≤50 mV peak-to-peak / ≤10 mV RMS
Voltage rise, 10–90% F.S., no load or full load≤3 s≤4 s≤8 s
Voltage fall, 90–10% F.S., no load≤15 s≤30 s≤60 s
Voltage fall, 90–10% F.S., full load≤3 s≤3 s≤3 s

F.S. denotes the respective full-scale value. Voltage line regulation and load regulation are each ≤0.01%. Resolution and accuracy are separate specifications. The longer no-load fall times must be considered when defining test and waiting times.

Operating modes and programmable functions

FunctionSettings and operation
SourceSet output voltage CV and current parameter I-Out; define V-Clamp and I-trip as trip thresholds.
Output ReverseSelect output polarity using SW or the Output Reverse setting.
SweepSet V-Start, V-End, V-Increment, Time Interval and current limit I-Out; select Single or Cont operation.
External ProgrammingEnable external voltage programming through VSET; the 0–10 V range corresponds to the voltage control range of the model.
PC recordingDisplay voltage and current, view traces, store readings and export data for further analysis.

Typical applications

Component breakdown testing

Apply a defined DC test voltage to high-voltage components such as IGBTs. Adjustable voltage levels and sweep steps support investigation of breakdown behaviour.

Insulation and material testing

Investigate insulating materials and components under high DC voltage. Model ranges up to 10 kV and low output currents provide the electrical basis for the test setup.

Leakage current and high resistance

Apply high voltage to a DUT and investigate its current response. Low ripple and voltage and current readback support sensitive measurements.

High-energy physics

Supply experimental setups with high DC voltage at low current. Selectable polarity and remote control support adaptation to individual research requirements.

Interfaces, software and automation

ConnectionAvailabilityFunction
LAN / RS232Built into the instrumentDigital remote control with SCPI support. The PC software connects over Ethernet.
VSETAnalogue voltage programming0–10 V; input impedance 1 MΩ; accuracy 0.2% F.S.; update rate 16 Hz.
VMON / IMONAnalogue voltage/current monitoringEach 0–10 V corresponding to the model range; output impedance 1 Ω; accuracy 0.2% F.S.; update rate 8 Hz.
PC softwareOperation and measurement analysisSource and sweep operation, graphical display, recording, historical data and Excel export.
NE101Optional, sold separatelyRS232-to-GPIB communication converter for integration into IEEE-488 systems.

Installation and infrastructure

ItemSpecification
AC input110/220 V AC ±10%, selected by a switch; 47–63 Hz.
Enclosure and dimensions2U, half 19-inch width; 88 × 214 × 413 mm (H × W × D).
WeightApproximately 3 kg.
EnvironmentOperation 0–40 °C; storage −20–60 °C; relative humidity 5–90%, non-condensing; altitude below 2,000 m.
Rack installationOptional bracket for a 19-inch rack, sold separately.
Connections and ventilationRear high-voltage output; provide reliable grounding and keep ventilation openings unobstructed.

Protection functions

OVP, OCP and OTP monitor overvoltage, overcurrent and overtemperature. In source mode, the output stops when the configured V-Clamp voltage threshold or I-trip current threshold is exceeded. The PC software also provides a configurable output-off response to a communication loss.

N3200 series FAQ

Which version reaches 10 kV?

The N3225-10K-M002 provides one channel covering −10,000 to 0 V or 0 to +10,000 V, with a maximum current of 2 mA and rated power of 20 W.

Can the output voltage be set to 0 V?

Yes. All three models can be adjusted down to 0 V in the selected positive or negative polarity.

What does ±2.5 kV, ±5 kV or ±10 kV mean?

It specifies the positive and negative voltage range of each model. Output Reverse selects the output direction; each instrument has one channel.

Can an external voltage control the output?

Yes. With external programming enabled, a 0–10 V signal at VSET controls output voltage. VMON and IMON separately monitor voltage and current.

Is GPIB built in?

The digital instrument interfaces are LAN and RS232. GPIB requires the optional NE101 converter, which is sold separately.

Can readings be saved?

The PC software records voltage and current values, displays traces and historical data, and supports export as an Excel file.

Can N3200 be used for IGBT breakdown testing?

Yes, breakdown testing of high-voltage components is an application. The required test voltage and current limit must remain within the selected model range.

Can the series be used for leakage current measurements?

Yes. It supplies the DC test voltage and provides current readback with 0.1 µA resolution. The required accuracy of the complete measurement setup must be assessed separately.

Are insulation and resistance investigations possible?

Yes. Insulation testing and high-resistance investigations are applications. The power supply provides the adjustable high voltage; test voltage, measurement range and setup must be matched to the DUT.

Technical consultation and project enquiry

Model selection depends on test voltage, polarity, maximum current, required resolution and measurement accuracy, and the intended test sequence. Please also specify your interfaces and whether you plan a benchtop or rack installation.

Discuss your N3200 test setup with ET System

Get in touch with us

+49 (0) 6205 - 3948-0 info@et-system.de

Mon - Thu: 8:00 AM - 5:00 PM
Fri: 8:00 AM - 3:00 PM

Contact Us Directly

+49 (0) 6205 - 3948-0 info@et-system.de

Mo - Do von 8 - 17 Uhr
Fr von 8 - 15 Uhr

Contact Us Directly